Click here to view the chromatograms and mass spectra for this data file in PDF format
Data File: /usr/users/finnigan/data/novatia/myolc02.dat
Sample: apomyoglobin 6.7uM x 20 ul inj
Comment: 20-90%B step in 20 min 1x10mm brownlee 75ul/min ACN/H2O/TFA
Client: unspecified
Deconvolution Parameters:
INMR= 499.988403 2000.032104
INTHRD= 0.02
OUTMR= 12750 21250
OUTMS= 1
PWIDTH= 3.0
MULTINT= 0.1
ADDION= 1.0079
Target Mass= 17000 +- 3.5
Spectrum Avg Width= 4


Input File: myolc02_sum1; Output File: myolc02_dec1
Scan: 213
Retention Time (min): 17.80
16949.9922 +- 1.8063
17004.8652 +- 1.6618


Input File: myolc02_sum2; Output File: myolc02_dec2
Scan: 235
Retention Time (min): 19.64
16951.0137 +- 2.7686
16980.8418 +- 0.8864


Input File: myolc02_sum3; Output File: myolc02_dec3
Scan: 245
Retention Time (min): 20.47
16951.0176 +- 2.5303
16995.5879 +- 2.4029


Input File: myolc02_sum4; Output File: myolc02_dec4
Scan: 247
Retention Time (min): 20.64
16950.8711 +- 2.3129
17003.5996 +- 2.5265


Input File: myolc02_sum5; Output File: myolc02_dec5
Scan: 294
Retention Time (min): 24.55


Target mass not found, however, a mass close to the target was found within 5% of the expected